• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS XCELA POWER INJECTABLE PICC; PERIPHERALLY INSERTED CENTRAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS XCELA POWER INJECTABLE PICC; PERIPHERALLY INSERTED CENTRAL CATHETER Back to Search Results
Catalog Number 60M701532
Medical Device Problem Code Material Discolored (1170)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/18/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported device has been returned to the manufacturer and an investigation into the root cause for event is currently in progress.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Event or Problem Description
An end user reported an issue with a can 5f dual xcela picc.It was reported that, during visual inspection, one lumen of the picc was noted to be cloudy yellow.There is also a "homogenous cloudiness" throughout the visible portion of the catheter.The picc was in situ for approximately 8 months before it was removed and replaced.The patient did not experience any adverse effects, harm, or require medical intervention because of this incident.The reported device has been returned to the manufacturer and an investigation into the root cause for event is currently in progress.
 
Additional Manufacturer Narrative
Returned for evaluation was a 5f dual lumen xcela pasv picc.As received, the customers complaint description is confirmed for lumen discoloration.The extension tubing had a frosted like appearance.Also noted the catheter tubing appeared pinkish in color.The extension tubing was sectioned, and it was observed that a sticky film was present on the inside surface that could be wiped off with isopropyl alcohol, leaving the extension tubing clear with no evidence of degradation.The composition of this film is unknown, however, based upon the absence of any damage to the extension tubing underneath the film, it appears to be a residue or precipitate from the infused drugs.Based on the sample analysis, this device did not malfunction.It is possible that the film observed on the id of the extension tubes was the result of instability of the etoposide mixture used/infused causing the drug to precipitate from solution as the observed film.The customer's reported complaint description of cloudy extension tubing is confirmed based on evaluation of returned picc complaint sample.The likely root cause for the discoloration is drug precipitate.This is an issue of customer use of the picc for a specific patient population drug therapy and not a picc device malfunction.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.The review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XCELA POWER INJECTABLE PICC
Common Device Name
PERIPHERALLY INSERTED CENTRAL CATHETER
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls, ny 12801 NY 12801
MDR Report Key12475954
Report Number1317056-2021-00165
Device Sequence Number3779729
Product Code LJS
UDI-Device IdentifierH96560M7015321
UDI-PublicH96560M7015321
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K070002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number60M701532
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/20/2021
Supplement Date Received by Manufacturer10/28/2021
Initial Report FDA Received Date09/15/2021
Supplement Report FDA Received Date10/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-