• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION Back to Search Results
Model Number MC2-5570S
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
 
Event or Problem Description
Information was received that the rod could not be distracted anymore.No patient adverse event was reported.
 
Additional Manufacturer Narrative
Updated section: the nail was received for visual inspection and functional testing.Upon return, visual inspection of the rod revealed evidence of discoloration.In addition, examination of the rod showed there was an attempt made to distract before it reportedly stopped to distract.This could happen if the rod was jammed if it was tried to retract back too much.Per reported failure, functional testing was not possible.X-ray images of the rod reveal that the retainer ring that should be position inside the groove is out of position.This could even prevent any rotation of magnet itself.This confirms that the rod is jammed which is the reason behind its failure to distract.A device history review (dhr) was performed on lot number 9021405 and there were no deviations in the manufacturing process and the finished product met all acceptance criteria prior to shipment.
 
Event or Problem Description
No additional information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Common Device Name
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, suite 100
aliso viejo CA 92656
MDR Report Key12477669
Report Number3006179046-2021-00458
Device Sequence Number14426040
Product Code PGN
UDI-Device Identifier00812258022648
UDI-Public812258022648
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K171791
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMC2-5570S
Device Lot Number9021405
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/02/2021
Supplement Date Received by Manufacturer10/26/2021
Initial Report FDA Received Date09/15/2021
Supplement Report FDA Received Date10/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-