Model Number 71430324
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/23/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference case (b)(4).
 
Event or Problem Description
It was reported that one gii art ins spcr trl 1-2 13mm was broken.It is unknown whether this problem was noticed in a surgical environment or not; therefore, patient involvement has not been confirmed.
 
Additional Manufacturer Narrative
H3, h6: the device, used in treatment, was returned for evaluation.A visual inspection of the returned trial confirm the stated failure.The device has multiple scratches, burrs, and deep gouges in the plastic.There is a piece of the device missing.The piece was not returned for evaluation.This device exhibits signs of extreme use and wear.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries and cleaning cycles.As plastics are vulnerable and crack may have initiated during use and possible causes could be due to the heating and cooling associated with autoclaving or prolonged use.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII ART INS SPCR TRL 1-2 13MM
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12477753
Report Number1020279-2021-06897
Device Sequence Number11600478
Product Code LXH
UDI-Device Identifier03596010481412
UDI-Public03596010481412
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71430324
Device Catalogue Number71430324
Device Lot Number12EB01776
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2021
Initial Date Received by Manufacturer 08/24/2021
Supplement Date Received by Manufacturer09/17/2021
Initial Report FDA Received Date09/15/2021
Supplement Report FDA Received Date09/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-