Model Number MC2-5570R
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
 
Event or Problem Description
Information was received that the rod could not be distracted anymore.No patient adverse event was reported.
 
Additional Manufacturer Narrative
The returned rod was observed to be partially distracted with some score marks on the distraction rod indicating the rod was partially distracted.The rod was functionally tested and was able to distract and retract with the manual distractor.The rod was able to retract/distract with fast distractor to perform stroke test the maximum and minimum length were measured at 303.43mm and 275.87mm respectively; giving a total stroke measurement of 27.56mm, which meets the 27-29mm specification.The work order was reviewed and confirmed the rod passed all inspections.Review of the device history records revealed no discrepancies related to this complaint.The rod was manufactured in accordance with the specified requirements and met all of the required quality inspections.
 
Event or Problem Description
No additional information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Common Device Name
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, suite 100
aliso viejo CA 92656
MDR Report Key12477918
Report Number3006179046-2021-00459
Device Sequence Number3892015
Product Code PGN
UDI-Device Identifier00812258022655
UDI-Public812258022655
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K171791
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMC2-5570R
Device Lot Number9022814
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/02/2021
Supplement Date Received by Manufacturer10/22/2021
Initial Report FDA Received Date09/15/2021
Supplement Report FDA Received Date10/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-