| Model Number |
801-05535D |
| Medical Device Problem Codes |
Fracture (1260); Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/17/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that 4 mesa foundation screws broke and the inner collets got "stuck into the rod" intra-operatively.Surgery was successfully completed with a 30 minute delay.No adverse consequences or medical intervention were reported.This record will capture the first of four screws.
|
| |
|
Event or Problem Description
|
|
It was reported that 4 mesa foundation screws broke and the inner collets got "stuck into the rod" intra-operatively.Surgery was successfully completed with a 30 minute delay.No adverse consequences or medical intervention were reported.This record will capture the first of four screws.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection: it was observed that the tabs on the inner collet had fractured.Deformation was observed on the outer and inner collet as well.Device and complaint history records were reviewed for this lot, and no relevant manufacturing issues or similar complaints were identified.It was communicated by the initial reporter that a mesa rod unlocker, instead of a mesa rail unlocker, was used.It is likely that the use of a mesa rod unlocker on a rail created misalignment on the screw head and excess loading during unlocking, resulting in the observed tab fracture.The cause of the reported issue will be classified as 'deviation from surgical technique'.
|
| |
|