• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL LOCKING SCREW, FULLY THREADED T2 HUMERUS Ø4X34 MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL LOCKING SCREW, FULLY THREADED T2 HUMERUS Ø4X34 MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 1896-4024S
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 08/26/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device has not been returned.If additional information becomes available, it will be provided in a supplemental report.Device remains implanted.
 
Event or Problem Description
It was reported a revision of the surgery performed about a year ago was scheduled due to confirmed postoperative screw fracture.However, before the revision, a mistake in ordering the instruments was discovered and the surgery was postponed.The patient was under general anesthesia.Revision surgery was rescheduled for (b)(6) 2021.This record captures the postoperative screw fracture.
 
Additional Manufacturer Narrative
The reported event could not be confirmed since the device was not returned for evaluation and no other evidence was provided.A device inspection was not possible since the affected device was not returned, and no other evidence were provided for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
It was reported a revision of the surgery performed about a year ago was scheduled due to confirmed postoperative screw fracture.However, before the revision, a mistake in ordering the instruments was discovered and the surgery was postponed.The patient was under general anesthesia.Revision surgery was rescheduled for (b)(6) 2021.This record captures the postoperative screw fracture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW, FULLY THREADED T2 HUMERUS Ø4X34 MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
marilyne chaumont
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key12478320
Report Number0009610622-2021-00692
Device Sequence Number4739469
Product Code HSB
UDI-Device Identifier04546540202307
UDI-Public04546540202307
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K003018
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/30/2024
Device Model Number1896-4024S
Device Catalogue Number18964034S
Device Lot NumberK0866D6
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/26/2021
Supplement Date Received by Manufacturer11/01/2021
Initial Report FDA Received Date09/15/2021
Supplement Report FDA Received Date11/29/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/02/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age89 YR
Patient SexFemale
Patient RaceAsian
-
-