MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. FIBEROPTIX ULTRA 8 IAB: 8FR 40CC; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. FIBEROPTIX ULTRA 8 IAB: 8FR 40CC; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN000254
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/12/2021
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the intra-aortic balloon (iab) ruptured.It was noted that a high-pressure alarm occurred, and blood was in the gas tubing.As a result, the iab was removed without difficulty and a second iab was inserted at a different insertion site.There was no report of patient complications or death.
 
Manufacturer Narrative
Qn#(b)(4).Teleflex received the device for investigation.The reported complaint of iab blood in helium pathway is confirmed.A puncture to the bladder, consistent with contact from the broken fiber, was found near the distal tip of the catheter which allowed blood to enter the helium pathway.The root cause of the broken fiber is undetermined.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.Teleflex assessed the risk for the reported complaint.There are no new or revised risk.This will be monitored for any developing trends.
 
Event Description
It was reported that the intra-aortic balloon (iab) ruptured.It was noted that a high-pressure alarm occurred, and blood was in the gas tubing.As a result, the iab was removed without difficulty and a second iab was inserted at a different insertion site.There was no report of patient complications or death.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIBEROPTIX ULTRA 8 IAB: 8FR 40CC
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key12480274
MDR Text Key271649631
Report Number3010532612-2021-00256
Device Sequence Number1
Product Code DSP
UDI-Device Identifier00801902007247
UDI-Public00801902007247
Combination Product (y/n)N
PMA/PMN Number
K021462
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/16/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2023
Device Model NumberIPN000254
Device Catalogue NumberIAB-05840-LWS
Device Lot Number18F21D0024
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/25/2021
Date Manufacturer Received10/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-