|
This complaint was reopened in reference to (b)(4).Investigation: the valve was received submersed in an unidentified liquid.The valve was set to pressure range 8 cmh2o at time of receipt.Visual inspection: no significant deformations or damage of the valve were detected during the visual inspection.Result: in the visual inspection of the valve, we could not detect any apparent damage.It was possible to adjust the valves up and down again in steps of 5 cmh2o and in steps of 2 cmh2o in the same way.To proof if the brake function is full in place we carried out a brake function test.The investigation has shown that the brake function is within the accepted tolerance.Further we carried out a permeability test.The investigation has resulted that the valve is permeable.As a final examination we carried out a computer controlled test.Here the progav 2.0 was tested in the lying position with a pressure range of 5 cmh2o.Normally we expected a pressure of 5 cmh2o having an opening pressure of 5 cmh2o.At the first test the opening pressure at the reference flow level of 5 ml/h is measured 1,12 cmh2o.The progav 2.0 valve tends to an over-drainage.We execute a second test.At the second test the opening pressure at the reference flow level of 5 ml/h is measured 0,44 cmh2o.The second test showed no improvement.The valve is not working inside the accepted tolerance (accepted tolerance of 5 ± 2 cmh2o).We suspect that deposits inside the valve may have affected the product performance.Given the fact that during the treatment with shunts the following complications may arise (as described in the literature): infections, blockages caused by protein and/or blood in the cerebrospinal fluid, over/under drainage or in very rare cases, noise development, we decided to dismantle the valve.Inside the valve we have found deposits or substances of protein, which probably could have caused the observed over-drainage.There was no failure detectable at the time of delivery.No further actions required.
|