| Model Number |
221165 |
| Medical Device Problem Code |
Contamination (1120)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/20/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that while using 20 bd bbl¿ columbia agars with 5% sheep blood fungal contamination was observed by the laboratory personnel.There was no indication that results were reported out and there was no report of patient impact.The following information was provided by the initial reporter: "they have received ship blood columbia agar plate.But when they open the box , they found fungus in the plate".
|
| |
|
Event or Problem Description
|
|
It was reported that while using 20 bd bbl¿ columbia agars with 5% sheep blood fungal contamination was observed by the laboratory personnel.There was no indication that results were reported out and there was no report of patient impact.The following information was provided by the initial reporter: "they have received ship blood columbia agar plate.But when they open the box , they found fungus in the plate".
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation: during manufacturing of material 221165, media is formulated and sent through a high temperature short time sterilizer to remove bioburden.The petri dishes are subjected to uv radiation to decrease bioburden.The petri dishes are filled in a positive pressure hepa filtered environment.The filled plates are cooled and immediately wrapped into sleeves to decrease the introduction of microbes.Sleeves are then packaged into cartons and then transferred to a refrigerated truck (2 to 8 degrees c) for shipment to the distributor.Bd distributors are provided with the storage guidelines for the shipping and handling of bd media of 2 to 8 degrees c in a dark place.The batch history record for batch 1134869 was satisfactory and no quality notifications were generated during manufacturing and inspection.The release testing that is performed on this product does include bioburden testing.A sample of plates are incubated at 25 degrees c and at 35 degrees c for approximately 72 hours.All bioburden testing performed on this batch was satisfactory per bd internal procedures.Affected product does not have any sterility claims; the product is tested for bioburden prior to release to ensure that it conforms to product specifications.However, this does not ensure that the end-user will not receive a contaminated plate.The complaint history was reviewed, and no other complaints have been taken on batch 1134869.Retention samples from batch 1134869 were not available for inspection.Two photos were received for investigation.One photo features a carton label from batch 1134865 (carton 0006) for batch verification.The other photo shows the agar surface of 8 plates with at least one microbial colony on the surface of each plate.No return samples were received for investigation.This complaint can be confirmed.A trend was identified for contamination and investigation found opportunities for bioburden reduction in the manufacturing process.Capa#3076308 has been initiated and involves implementing additional cleaning events and evaluation of manufacturing procedures focused on in-process bioburden reduction.Additional trainings are planned with an ongoing training review for cleaning processes.Bd will continue to trend complaints for contamination.
|
| |
|
Search Alerts/Recalls
|