| Model Number |
CD3265-40 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/30/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device is included in the battery performance alert advisory issued by abbott on 28 august 2017.
|
| |
|
Event or Problem Description
|
|
Following the battery performance alert (bpa) advisory, a bpa was received by the clinician and the device was explanted.The patient was asymptomatic and in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
Bpa was not confirmed by analysis.Interrogation of the device revealed the device was above eri when received.A longevity calculation was performed and found the battery depletion was normal based on the device usage.Based on this information, there was no evidence of premature depletion.
|
| |
|