CONMED CORPORATION RAPIDVAC; APPARATUS, EXHAUST, SURGICAL Back to Search Results
Model Number SE3695
Medical Device Problem Codes Communication or Transmission Problem (2896); Electrical Shorting (2926)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/31/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during use, safety test failed due to excessive leakage current and the keypad was damaged.There was no patient injury.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6 correction: d3(mfr name, street 1 and mfr city) h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the unit had a motor and keypad failure.It was reported that the device had current leakage and the button on the device had a failure.The reported issues were confirmed.The most likely cause was traced to a component failure.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDVAC
Common Device Name
APPARATUS, EXHAUST, SURGICAL
Manufacturer (Section D)
CONMED CORPORATION
6455 s yosemite st ste 800
greenwood village CO 80301 3299
MDR Report Key12484911
Report Number1717344-2021-01280
Device Sequence Number3936974
Product Code FYD
UDI-Device Identifier10884524001753
UDI-Public10884524001753
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K980915
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSE3695
Device Catalogue NumberSE3695
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2021
Initial Date Received by Manufacturer 09/01/2021
Supplement Date Received by Manufacturer09/20/2021
Initial Report FDA Received Date09/16/2021
Supplement Report FDA Received Date10/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-