• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK SURPASS EVOLVE 4.0MM X 20MM - PMAS; INTRACRANIAL ANEURYSM FLOW DIVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK SURPASS EVOLVE 4.0MM X 20MM - PMAS; INTRACRANIAL ANEURYSM FLOW DIVERTER Back to Search Results
Model Number FDS40020
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Headache (1880); Stenosis (2263)
Date of Event 06/02/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device is implanted in patient.
 
Event or Problem Description
It was reported that during a clinical trial procedure for a left ophthalmic saccular aneurysm in the internal carotid artery, the subject flow diverter was implanted successfully.Pre procedure imaging did not indicate any additional abnormalities.Procedure angiogram indicated 0-25% stenosis in the parent artery and in the subject flow diverter implanted which had no clinical significance to the patient.One day post procedure, the patient experienced non serious persistent ipsilateral headaches that were unrelieved by tylenol.Patient medication was changed to fioricet and the headaches resolved.No other information is available.
 
Event or Problem Description
It was reported that during a clinical trial procedure for a left ophthalmic saccular aneurysm in the internal carotid artery, the subject flow diverter was implanted successfully.Pre procedure imaging did not indicate any additional abnormalities.Procedure angiogram indicated 0-25% stenosis in the parent artery and in the subject flow diverter implanted which had no clinical significance to the patient.One day post procedure, the patient experienced non serious persistent ipsilateral headaches that were unrelieved by tylenol.Patient medication was changed to fioricet and the headaches resolved.No other information is available.
 
Additional Manufacturer Narrative
Based on the results of the dhr review, there is no indication that the device, labeling or packaging failed to meet its specifications when released.The reported complaint not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.Per the ae description ¿persistent left sided headaches unrelieved by tylenol no visual changes started fioricet¿.This is classified as a worsening focal ipsilateral headache by the site but not a suspected stoke event.Medication was given for the event, no other medical intervention is indicated in edc.Procedure angiogram indicates 0-25% stenosis (which is equivalent to no clinically significant stenosis).The reported 'patient parent vessel stenosis' and 'patient headache non-serious', are known and anticipated complications to these types of procedures and patient condition and is listed as such in the device directions for use.Therefore, a probable cause of anticipated procedural complication was assigned to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURPASS EVOLVE 4.0MM X 20MM - PMAS
Common Device Name
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key12485706
Report Number3008881809-2021-00386
Device Sequence Number20105622
Product Code OUT
UDI-Device Identifier07613327375206
UDI-Public07613327375206
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P170024/S003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/05/2023
Device Model NumberFDS40020
Device Catalogue NumberFDS40020
Device Lot Number22510830
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/23/2021
Supplement Date Received by Manufacturer12/20/2021
Initial Report FDA Received Date09/16/2021
Supplement Report FDA Received Date01/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/06/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age41 YR
Patient SexFemale
Patient Weight62 KG
Patient RaceWhite
-
-