U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US LCS COMPLETE FEM CEM R LG; LCS COMPLETE : KNEE FEMORAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US LCS COMPLETE FEM CEM R LG; LCS COMPLETE : KNEE FEMORAL Back to Search Results
Model Number 1294-01-060
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/26/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The patient was revised to address loosening of the femoral and tibial components at the bone to cement and cement to implant interfaces.Unknown bone cement was used.Osteolysis was also noted.Doi: 2000, dor: (b)(6) 2021, right knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.No product images or patient x-rays were provided.Examination did not find sufficient evidence to confirm loosening of the reported femoral insert.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LCS COMPLETE FEM CEM R LG
Common Device Name
LCS COMPLETE : KNEE FEMORAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key12485954
Report Number1818910-2021-20347
Device Sequence Number14290251
Product Code NJL
UDI-Device Identifier10603295021575
UDI-Public10603295021575
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2000
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1294-01-060
Device Catalogue Number129401060
Device Lot Number2201058
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/26/2021
Supplement Date Received by Manufacturer10/25/2021
Initial Report FDA Received Date09/16/2021
Supplement Report FDA Received Date10/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LARGE 3 PEG PATELLA POLY; LCS COMP RP INSERT LG 17.5MM; LCS COMPLETE FEM CEM R LG; MBT CEM TIBIAL TRAY SZ5; UNK MANUFACTURER & PRODUCT INFO; LARGE 3 PEG PATELLA POLY; LCS COMP RP INSERT LG 17.5MM; LCS COMPLETE FEM CEM R LG; MBT CEM TIBIAL TRAY SZ5; UNK MANUFACTURER & PRODUCT INFO
Outcome Attributed to Adverse Event Required Intervention;
Patient Age78 YR
-
-