| Model Number |
1220-72-332 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Foreign Body Reaction (1868); Pain (1994); Joint Dislocation (2374); Osteolysis (2377); Metal Related Pathology (4530); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
01/01/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Medical records received.After review of medical records, it was reported that the patient presented in er due to severe pain and dislocation.Closed reduction was performed.Mri reported of osteolysis associated with metallosis.Doi: (b)(6), 2017.Dor: none reported.Left hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4) investigation summary : no device associated with this report was received for examination.All available x-rays were reviewed and the complaint cannot be confirmed.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H6 health effect - clinical code: appropriate term / code not available (e2402) used to capture the blood heavy metal increased.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.All available x-rays were reviewed and the complaint cannot be confirmed.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Medical records were received: on (b)(6) /2017, the patient had left pain and clicking, an mr of the left hip was reported to show a small hip effusion, synovial thickening within the iliopsoas bursa, along with right trochanteric bursitis.(no mention of any invasive procedure) on (b)(6) 2017, the patient had a left total hip arthroplasty to address polyethylene wear.A preoperative radiograph was reported to show some polyethylene wear.On (b)(6) 2018 the patient has a medical diagnosis of tha revision, and dislocation (no mention of when that dislocation had happened).
|
| |
|
Search Alerts/Recalls
|