• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PFNA BLADE PERF L100 TAN; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PFNA BLADE PERF L100 TAN; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 04.027.035S
Device Problem Device-Device Incompatibility (2919)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/02/2021
Event Type  malfunction  
Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Device is not distributed in the united states, but is similar to device marketed in the usa.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from (b)(6) reports an event as follows: it was reported that during a procedure on (b)(6) 2021, the pfna blade was in locked position when taken from the sterile package, prior to implantation in a patient.Another device had to be used.This report is for a pfna blade perf l100 tan.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H4, h6: product code: 04.027.035s, lot: 164p072, manufacturing site: bettlach, release to warehouse date: may 10, 2021, expiry date: may 01, 2031.A manufacturing record evaluation was performed for the finished device lot and no non-conformance was identified.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFNA BLADE PERF L100 TAN
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
MDR Report Key12487978
MDR Text Key271896489
Report Number8030965-2021-07830
Device Sequence Number1
Product Code HSB
UDI-Device Identifier07611819368996
UDI-Public(01)07611819368996
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/17/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number04.027.035S
Device Lot Number164P072
Was Device Available for Evaluation? No
Date Manufacturer Received10/08/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/10/2021
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
UNK - SCREWS: NAIL DISTAL LOCKING; UNK - SCREWS: NAIL DISTAL LOCKING
-
-