• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GNS II BICONVEX PAT TRIAL 26MM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GNS II BICONVEX PAT TRIAL 26MM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 71430568
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/27/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that, the listed devices were cracked or broken.As this was noticed upon field inspection, there was not patient involvement.
 
Additional Manufacturer Narrative
The device, intended for use in treatment, was not returned for evaluation and the reported event could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GNS II BICONVEX PAT TRIAL 26MM
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12488128
Report Number1020279-2021-06952
Device Sequence Number3391648
Product Code LXH
UDI-Device Identifier03596010212184
UDI-Public03596010212184
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71430568
Device Catalogue Number71430568
Device Lot Number18LB00874
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/27/2021
Supplement Date Received by Manufacturer10/28/2021
Initial Report FDA Received Date09/17/2021
Supplement Report FDA Received Date10/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-