Complainant alleged that while attempting to defibrillate a (b)(6) year-old male patient, the device was unable to obtain an ecg signal via electrode pads.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that the patient subsequently expired, however it was not a result of the reported malfunction.
|
The device was returned to zoll medical united kingdom.The customer's report was not replicated or confirmed.The device passed all testing without duplicating the report.Review of the device log shows that electrodes were applied.It takes several minutes to acquire an ecg signal with some troubleshooting.A second set of electrodes was most likely utilized.The ecg signal is present for 90 consecutive minutes and was a clean, strong signal.Without the return of the electrodes used, it cannot be determined if there was a concern with the electrodes, application of the electrodes, patient preparation, or other factor that would prevent or inhibit the acquisition of an ecg signal.The electrodes used were not retained and were unavailable for evaluation.Analysis of reports of this type has not identified an increase in trend.
|