The reported event for no ipg communication was confirmed.As received, the ipg was running the therapy application software.However, a review of the ipg reset log confirmed the device had a msp430 reset that occurred on (b)(6) 2021, the day of the implant procedure.Per event details, the ipg no longer communicated during the ipg implant procedure that used electrocautery.Despite the ipg having a msp430 reset, the device never entered the service application state and remained in the therapy application state.The msp430 reset would have temporarily caused an inability to communicate with a cp, which was observed in the field.
|