ROOT WITH NONINVASIVE BLOOD PRESSURE AND TEMPERATU; MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS) Back to Search Results
Model Number 26224
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/23/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
 
Event or Problem Description
The customer reported the blood pressure measurement is changing from extreme high to extreme low.No patient impact or consequences were reported.
 
Event or Problem Description
The customer reported the blood pressure measurement is changing from extreme high to extreme low.No patient impact or consequences were reported.
 
Additional Manufacturer Narrative
Additional manufacturing narrative: the returned device was evaluated.During evaluation the device passed all visual and functional testing.During simulation testing, the device passed manual and preset conditions and provided accurate measurements.The unit was found to visually and audibly alarm during alarm conditions.The unit was determined to be functioning as designed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROOT WITH NONINVASIVE BLOOD PRESSURE AND TEMPERATU
Common Device Name
MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS)
MDR Report Key12492442
Report Number3011353843-2021-00244
Device Sequence Number14792481
Product Code MWI
UDI-Device Identifier00843997011907
UDI-Public00843997011907
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K110028
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 08/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number26224
Device Catalogue Number9695
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/25/2021
Supplement Date Received by Manufacturer10/11/2021
Initial Report FDA Received Date09/17/2021
Supplement Report FDA Received Date11/04/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/10/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
-
-