| Model Number |
A22041T |
| Medical Device Problem Codes |
Break (1069); Fracture (1260); Material Fragmentation (1261); Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/15/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
|
| |
|
Event or Problem Description
|
|
Olympus was informed that during a therapeutic transurethral resection procedure it was noticed that the ceramic insulation at the distal end of the resection sheath was missing.The missing fragment could not be located and it is unclear whether it was already missing during the pre-use inspection or whether the fragment broke off inside the patient.The intended procedure was completed with another similar device and there was no report about an adverse event or patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect medical device was not returned to the manufacturer for investigation/evaluation but to olympus medical systems corporation (omsc), japan (returned to the omsc on 2021-09-27).The evaluation at the omsc confirmed that the ceramic insulation at the distal end of the resection sheath is broken off.The type of damage found is typically caused by thermal and/or mechanical fatigue caused by wear and tear, possibly in combination with excessive force.Therefore, this event/incident was attributed to use error.It cannot be conclusively determined whether the insulating insert was pre-damaged at some point in the past, whether the damage was triggered during the reprocessing cycle preceding the incident, or during the procedure.A material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the resection sheath without showing any abnormalities.The case will be closed on olympus side with no further actions but the reported event/incident will be recorded for trending and surveillance purposes.Furthermore, the user will be informed about the investigation results.
|
| |
|