MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH RESECTION SHEATH, 24 FR., WITH DEFLECTING OBTURATOR; RESECTION SHEATHS, IRRIGATION RINGS, CYSTOSCOPE SHEATHS, HYSTEROSCOPE SHEATHS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH RESECTION SHEATH, 24 FR., WITH DEFLECTING OBTURATOR; RESECTION SHEATHS, IRRIGATION RINGS, CYSTOSCOPE SHEATHS, HYSTEROSCOPE SHEATHS Back to Search Results
Model Number A22041T
Medical Device Problem Codes Break (1069); Fracture (1260); Material Fragmentation (1261); Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/15/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event or Problem Description
Olympus was informed that during a therapeutic transurethral resection procedure it was noticed that the ceramic insulation at the distal end of the resection sheath was missing.The missing fragment could not be located and it is unclear whether it was already missing during the pre-use inspection or whether the fragment broke off inside the patient.The intended procedure was completed with another similar device and there was no report about an adverse event or patient injury.
 
Additional Manufacturer Narrative
The suspect medical device was not returned to the manufacturer for investigation/evaluation but to olympus medical systems corporation (omsc), japan (returned to the omsc on 2021-09-27).The evaluation at the omsc confirmed that the ceramic insulation at the distal end of the resection sheath is broken off.The type of damage found is typically caused by thermal and/or mechanical fatigue caused by wear and tear, possibly in combination with excessive force.Therefore, this event/incident was attributed to use error.It cannot be conclusively determined whether the insulating insert was pre-damaged at some point in the past, whether the damage was triggered during the reprocessing cycle preceding the incident, or during the procedure.A material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the resection sheath without showing any abnormalities.The case will be closed on olympus side with no further actions but the reported event/incident will be recorded for trending and surveillance purposes.Furthermore, the user will be informed about the investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESECTION SHEATH, 24 FR., WITH DEFLECTING OBTURATOR
Common Device Name
RESECTION SHEATHS, IRRIGATION RINGS, CYSTOSCOPE SHEATHS, HYSTEROSCOPE SHEATHS
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
MDR Report Key12493931
Report Number9610773-2021-00257
Device Sequence Number19850739
Product Code FJL
UDI-Device Identifier04042761020978
UDI-Public04042761020978
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K931995
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 10/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA22041T
Device Catalogue NumberA22041T
Device Lot Number20401
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/15/2021
Supplement Date Received by Manufacturer10/13/2021
Initial Report FDA Received Date09/20/2021
Supplement Report FDA Received Date10/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OLYMPUS ELECTROSURGICAL GENERATOR "ESG-400"
-
-