Catalog Number UNKNOWN
Device Problems Structural Problem (2506); Device Tipped Over (2589)
Patient Problems Abdominal Pain (1685); Pain (1994); Perforation of Vessels (2135)
Event Type  Injury  
Manufacturer Narrative
Occupation: non-healthcare professional.Investigation: the reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: vena cava perforation, tilt.Filter interacts with ivc wall, e.G.Penetration/ perforation/ embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/ perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Catalog number and lot number are unknown, however, the alleged tulip is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.A follow-up medwatch report will be submitted if additional relevant information becomes available.
 
Event Description
The following information is alleged: the patient received a gunther tulip inferior vena cava (ivc) filter on (b)(6) 2008 due to deep vein thrombosis (dvt).Approximately 11 years and 2 months later the patient had a abdomen and pelvis computed tomography scan to evaluate the possibility of a dvt.The scan revealed tilt and perforation.Hospital and medical records have been requested, but not yet provided.
 
Event Description
Patient allegedly received a filter via the right common femoral vein after deep vein thrombosis (dvt) and pulmonary embolism (pe).Patient is alleging device tilt and vena cava perforation.Patient notes and further alleges experiencing "back, abdominal, and hip pain".Per a computed tomography (ct) abdomen: "findings: filter type: gunther tulip.Filter placement site: infrarenal ivc.Evidence of filter fracture: no.Evidence of filter deformation: no.Migration of filter after deployment: no, tip at inferior endplate of l1, unchanged compared to 8/7/08.Filter tilt: mildly limited due to convex left scoliosis, but unchanged and approximately <3 degrees.Did any filter legs appear outside the ivc? no (legs extend <3 mm from caval wall).Medial leg abuts but does not appear to invade the posterior, left lateral aspect of the aorta.Impression: unremarkable gunther tulip ivc filter in good position".
 
Manufacturer Narrative
Investigation: investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: abdominal/back/hip pain.Unknown if the reported abdominal, back, and hip pain are directly related to the filter and unable to identify a corresponding failure mode at this point in time.Catalog and lot numbers are unknown, however, the alleged tulip is manufactured and inspected according to controls.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Manufacturer Narrative
Investigation: investigation is reopened due to additional information provided.Per quality engineering review, the additional information provided for this complaint does not change the previous investigation conclusion.Therefore, no new investigation activities will be conducted at this time.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
Per an undated medical opinion of medical imaging (various dates): "positive for perforation (grade 2) and tilt, negative for migration".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
larry pool
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key12495056
MDR Text Key272140931
Report Number1820334-2021-02174
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 08/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/25/2021
Initial Date FDA Received09/20/2021
Supplement Dates Manufacturer Received11/22/2021
07/21/2022
Supplement Dates FDA Received12/14/2021
08/10/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient SexMale
Patient Weight70 KG
-
-