H11: b5- the manufacturer was contacted in reference to the voluntary field safety notice/ recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously received information alleging an issue related to cpap/bipap device's sound abatement foam.The patient alleged that the visualization of particles.The patient alleged nasal allergies.H10: additional information was received on 09/27/2023.The device was returned to the manufacturer's service center for further evaluation.There was no mention of visual findings to the external part of the device.The internal aspect of the device was inspected.The device powered on, and the airflow was confirmed.The device's downloaded logs were reviewed.The manufacturer concludes that the customer allegation could not be confirmed and there was no visible foam degradation.The unit has been scrapped.Sections d8, d9, h3 and h6 updated.
|