DEPUY ORTHOPAEDICS INC US TRI-LOCK BPS SZ 4 STD OFFSET; TRILOCK HIP STEM : HIP FEMORAL STEM Back to Search Results
Model Number 1012-04-040
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Codes Pain (1994); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 01/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that in (b)(6) 2018, the patient fell from a truck bed and fractured his femoral neck and underwent the bha surgery with the stem.The date of fall and the surgery is unknown.After the surgery, the patient complained of pain during hospitalization, but was discharged from the hospital.After that, although the patient was followed regularly, but the occurrence of migration of the stem had not been noticed, and the surgeon found the migration of the stem at the latest medical examination.X-ray showed that migration had been considerably progressing from 1 year ago.The surgeon said that migration might have progressed quickly because the patient was still very active, such as loading and unloading trucks, and he is still in his 50s.In addition, the bone on the greater trochanter side of the stem has dissolved, and infection is suspected.The revision surgery via tha is scheduled on (b)(6) 2021.
 
Additional Manufacturer Narrative
Product complaint # : (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-LOCK BPS SZ 4 STD OFFSET
Common Device Name
TRILOCK HIP STEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key12495501
Report Number1818910-2021-20468
Device Sequence Number14430392
Product Code KWA
UDI-Device Identifier10603295000754
UDI-Public10603295000754
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P070026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1012-04-040
Device Catalogue Number101204040
Device Lot NumberHP3648
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/27/2021
Supplement Date Received by Manufacturer10/13/2021
Initial Report FDA Received Date09/20/2021
Supplement Report FDA Received Date10/13/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARTICUL/EZE BALL 28 +5 BR; J-FX BIPOL SHELL&28MMLINR/51MM; TRI-LOCK BPS SZ 4 STD OFFSET
-
-