MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 AXIUM 3D; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 AXIUM 3D; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number QC-2.5-4-3D
Medical Device Problem Codes Therapy Delivered to Incorrect Body Area (1508); Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/15/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that a axium coil would not detach with the use of an instant detacher.The physician then reverted to breaking the hypotube and manually pulling the detachment filament and it still wouldn't detach.The coil was being removed from the patient the coil detached in the microcatheter.The microcatheter then had to be removed and replaced.The physician was using axium coils to treat a small ruptured acom aneurysm.The aneurysm was saccular with a small neck and max diameter.Patient blood flow was normal. the case then continued with the successful deployment of more axium coils.The physician did not try to detach with another instant detacher.There was one attempt to detach via the hypotube.No symptoms were reported.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information received reported that the same instant detacher was used with other coils in the case without any problems.Prior to the non-detachment there had been no issues, and no damage was observed to the pushwire after it was removed from the patient.
 
Additional Manufacturer Narrative
Product analysis: ¿ as found condition: the axium coil and pushwire were returned for evaluation within the outer carton; inside of a biohazard bag and a shipping box.¿ visual inspection/damage location details: the implant coil appeared to be detached from the pushwire.The implant coil was found to be stretched.The shield coil was found intact with no damage.The coin was not located against the lumen stop as it was pulling back.The pusher was found broken at the break indicator (manual detachment location); with the release wire pulling back; it appeared that the manual detachment was attempted at this location.The ai and coupler tubing were found missing from the pushwire and not returned.¿ testing/analysis: under the microscope, the outer jacket was then removed to gain access the coin.The coin was measured in 3 locations (0.077mm @ 0.063mm; measured 0.086mm @ 0.127mm; measured 0.099mm @ 0.275mm) and found to be within specifications.The inner diameter of the lumen stop, and the inner diameter of the retainer ring were found to be visually acceptable.The lumen stop inner diameter (id) was measured to be 0.00275¿ and found to be within specification.The retainer ring inner diameter (id) was measured to be 0.00462¿and found to be within specification.The detach element was in good shape and the detachment ball was measured to be 0.0038" which was found to be within specifications.All other subassemblies appeared to be normal, and no other anomalies were observed.¿ conclusion: based on the analysis performed, the axium coil was not confirmed to have "non-detachment" issue as the pushwire was returned with the implant coil already detached.Based on the returned device, there was evidence of manual detachment attempt to detach the coil as the pusher was also found to be broken at the manual detachment location; with the coin pulled back from the lumen stop.The pulling back of the coin from the lumen stop may have contributed to the detachment of the implant coil from the pushwire.There was no non-conformance to specification identified that that led to the non-detachment issue.The review of lot history records shows that the finished device has met all manufacturing requirements and specifications during final assembly and quality inspection.Since the ai and coupler tubing were not returned; any contribution of the ai and coupler tubing to the reported issue could not be determined.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIUM 3D
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key12496732
Report Number9612164-2021-03602
Device Sequence Number3398029
Product Code KRD
UDI-Device Identifier00847536037656
UDI-Public00847536037656
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K081465
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/17/2024
Device Model NumberQC-2.5-4-3D
Device Catalogue NumberQC-2.5-4-3D
Device Lot Number221730358
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/29/2021
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/17/2021
Supplement Date Received by Manufacturer11/03/2021
11/24/2021
Initial Report FDA Received Date09/20/2021
Supplement Report FDA Received Date11/07/2021
12/08/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-