|
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one dorado pta dilatation catheter has been received for the evaluation.On the visual evaluation, the device was noted to be bloody and no other specific anomalies were noted to the device.On the functional evaluation, an inhouse presto inflation device was used to inflate the sample and water was noted to be leaking from the glue joint.Under the microscopic observation, the circumferential break was noted at the proximal balloon joint.No other functional testing performed.Therefore the investigation for the reported inflation issue was confirmed, as the water starts leaking from the circumferential break at the proximal balloon joint upon inflation.The investigation for the identified balloon joint break was also confirmed, as the circumferential break was noted at the proximal balloon joint under the microscopic observation.The identified balloon joint break is likely the root cause for the reported inflation issue.However, the definitive root cause for the reported inflation issue and the identified balloon joint break could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 10/2023), h11: h6 (result, conclusion), h11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|