• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZEPHYR ENDOBRONCHIAL DELIVERY CATHETER; ENDOBRONCHIAL VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZEPHYR ENDOBRONCHIAL DELIVERY CATHETER; ENDOBRONCHIAL VALVE Back to Search Results
Model Number ZEPHYR 4.0 EDC
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2021
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2021, the patient underwent a bronchoscopic lung volume reduction (blvr) procedure with zephyr valves.The physician had used the zephyr 4.0 endobronchial delivery catheter to size airways in the left upper lobe (lul) and deliver valves.A small piece of the sizing wing from the catheter broke off and was removed from the patient with graspers.The physician used a new catheter and was able to successfully complete the procedure.The catheter was set aside to be returned to the manufacturer for evaluation but was discarded by a nurse in the room.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZEPHYR ENDOBRONCHIAL DELIVERY CATHETER
Common Device Name
ENDOBRONCHIAL VALVE
MDR Report Key12498080
Report Number3007797756-2021-00259
Device Sequence Number17090695
Product Code NJK
UDI-Device Identifier00811907030522
UDI-Public(01)00811907030522(10)505362V50(17)230120
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P180002
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/20/2023
Device Model NumberZEPHYR 4.0 EDC
Device Catalogue NumberEDC-TS-4.0
Device Lot Number505362-V5.0
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2021
Initial Report FDA Received Date09/20/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-