• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION X SERIES; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION X SERIES; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number X SERIES
Device Problem Unable to Obtain Readings (1516)
Patient Problem Necrosis (1971)
Event Date 07/31/2021
Event Type  Death  
Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Event Description
Complainant alleged that while attempting to defibrillate a (b)(6) year-old male patient, the device was unable to detect the attached electrode pads.Complainant indicated that the patient subsequently expired.
 
Manufacturer Narrative
The device was not returned to zoll medical corporation for evaluation.Instead, the clinical file of the customer's report was provided.Review of the clinical file shows that the user entered pacer mode, which will prevent the selection and viewing of the pads lead.This is by design and is normal operation of the device.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X SERIES
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key12498516
MDR Text Key272247198
Report Number1220908-2021-03021
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00847946005993
UDI-Public00847946005993
Combination Product (y/n)N
PMA/PMN Number
K112432/P160
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Type of Report Initial,Followup
Report Date 08/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/20/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Model NumberX SERIES
Device Catalogue NumberX SERIES
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age60 YR
-
-