| Model Number |
DJ4011X |
| Medical Device Problem Code |
Device Markings/Labelling Problem (2911)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This is a report about missing legally required label on package.In response to the occurrence reported in (b)(4), the distributor checked the products in the shelf carton in stock and found that there was no legally required label on the package of all the remaining eight products.The legally required label has been affixed to the shelf carton.Bdj is legally required as an mah to affix the label not on the package but on the shelf carton.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) initial emdr submission.A follow up emdr will be submitted if additional information becomes available.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Pr (b)(4) follow-up emdr for device evaluation: three photos sample were received by our quality team for investigation.Upon visual inspection of the photo sample, it was observed that the sealed package of jamshidi contains the needle, the extractor and the luer lock adaptor.White label is attached on top of the product information label.The label indicated in the description of the complaint is not required to be affixed to the units at the manufacturing plant.A review of the internal manufacturing device records and raw material history files for the reported lot number 0001329661 was performed and no recorded quality problems or rejections related to this incident were found.Product undergoes inspections throughout manufacturing, no issues related to the reported incident were identified, all procedural and functional requirements for product release have been met.The missing labels are attached at the dc not at the manufacturing plant.Based on the quality team's investigation, a probable root cause for the reported failure mode could not be established.Complaints received for this device and defect will continue to be monitored by our quality team for signs of emerging trends.H3 other text : see manufacturer here.
|
| |
|
Event or Problem Description
|
|
This is a report about missing legally required label on package.In response to the occurrence reported in (b)(4), the distributor checked the products in the shelf carton in stock and found that there was no legally required label on the package of all the remaining eight products.The legally required label has been affixed to the shelf carton.Bdj is legally required as an mah to affix the label not on the package but on the shelf carton.Per follow-up email: bdj received 8 unopened products and confirmed that the japanese labels were not affixed to the individual package.Please refer to attached photo.The actual sample won't be sent because it is a defect due to domestic work.
|
| |
|