| Model Number |
367987 |
| Medical Device Problem Codes |
Short Fill (1575); Volume Accuracy Problem (1675)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported when using the bd vacutainer® sst¿ blood collection tubes, the device experienced underfill or low draw of a tube with blood.The following information was provided by the initial reporter.The customer stated: it is reported customer is experiencing under filling.Tiger top sst 7.5ml, many of the tubes do not have a vacuum in them, seems to be random in that lot.
|
| |
|
Event or Problem Description
|
|
It was reported when using the bd vacutainer® sst¿ blood collection tubes, the device experienced underfill or low draw of a tube with blood.The following information was provided by the initial reporter.The customer stated: it is reported customer is experiencing under filling.Tiger top sst 7.5ml, many of the tubes do not have a vacuum in them, seems to be random in that lot.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation: bd had not received samples or photos for investigation.Therefore, thirty (30) retention samples from bd inventory were evaluated by functional testing and no issues were observed relating to underfill as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode of underfill.Bd was not able to identify a root cause for the indicated failure mode.
|
| |
|
Search Alerts/Recalls
|