• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER DENTAL IMPL TAPERED SCR-V HA 3.7 MM 3.5MM 16MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL IMPL TAPERED SCR-V HA 3.7 MM 3.5MM 16MM; DENTAL IMPLANT Back to Search Results
Model Number TSVH16
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/19/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet complaint number cmp-(b)(4).Age and date of birth unknown / not provided.Patient sex unknown / not provided.Weight unknown / not provided.Email address unknown / not provided.Additional pma/510(k) number ¿ k011028|k013227.A summary investigation has been completed for lack of primary stability events that do not allege a deficiency with the implant and identified that the reported event is likely due to biological factors which have an adverse effect on implant stability or is related to surgical technique and insertion torque.Due to a wide range of external factors (non-design/ non-manufacturing related), identifying a definitive root cause is generally not possible.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
It was reported that the implant failed to integrate due to lack of primary stability and was removed.The site was grafted.Tooth location unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPL TAPERED SCR-V HA 3.7 MM 3.5MM 16MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
MDR Report Key12502373
Report Number0002023141-2021-02606
Device Sequence Number3695647
Product Code DZE
UDI-Device Identifier00889024019591
UDI-Public(01)00889024019591(17)250609(10)1235500
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K061410
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/09/2025
Device Model NumberTSVH16
Device Catalogue NumberTSVH16
Device Lot Number1235500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/26/2021
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2021
Initial Report FDA Received Date09/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-