| Model Number |
TV-IL1412100-J |
| Medical Device Problem Codes |
Complete Blockage (1094); Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
08/26/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device remains implanted in the patient; therefore, it will not be returned for evaluation.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The patient was initially implanted with an afx2 bifurcated stent graft, a vela suprarenal extension and an ovation limb extension to treat an abdominal aortic aneurysm (aaa).This case is outside the indications of use (off -label) due the use of afx with ovation devices.Approximately one (1) day post initial procedure, the patient presented with thrombosis of the right limb causing an occlusion of the stent graft.The physician successfully treated the patient by implanting two (2) bare metal peripheral stents (non-endologix) and the patient is reportedly doing well.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.There are no other equivalent adverse events/incidents for this lot number existing within the endologix complaint handling system.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it was discarded by the facility.A clinical evaluation of the adverse event/incident could not be completed.No medical records nor medical imaging relevant to the reported adverse event/incident was received by endologix.The user facility was unable to provide this information.Due to the absence of medical records and medical imaging; device, use, procedure, and/or anatomy relatedness to this adverse event/incident could not be evaluated.During the investigation, endologix found reasonable evidence to suggest the device was used off-label.The reported concomitant product use of an ovation limb in an afx main body is off label.The final patient status was reported as doing well no additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.H6: result code: remove code 3233.H6: conclusion code: remove code 11.
|
| |
|
Search Alerts/Recalls
|