| Model Number |
DXT5JR40 |
| Medical Device Problem Codes |
Deformation Due to Compressive Stress (2889); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/17/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: one 5fr dexterity catheter was received for analysis.The catheter was kinked, flattened, and evidence of a torsional kink on the shaft of the catheter approx.46.2cm distal to the strain relief.The od was measured and confirmed to be 0.067inch with the spec being 0.070 max.The id was measured and confirmed to be 0.044 with spec being 0.042 min.No damage was noted to the distal tip.No other damage was noted to the remainder of the device.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
During a procedure an attempt was made to use two dexterity catheters to treat a lesion in the rca and lm arteries.There was no damage noted to either device.Devices were removed from the packaging per ifu.There was no resistance noted while removing the devices from packaging.A brachial approach was used for the procedure.The devices were inspected with issues noted.The first device was prepped per ifu.It was reported that this device was kinked in the middle of the catheter.The kink occurred while the device was in the patient.Excessive force was not used and the device was not excessively torqued.Resistance was not experienced.It was reported that the second device seemed to be kinked.The kink was noted on the device shaft while torquing the catheter to engage the artery.The device was excessively torqued.Resistance was not experienced and excessive force was not used.The patient is alive with no injury.
|
| |
|
Search Alerts/Recalls
|