• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH MRH KNEE FEM S LFT; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH MRH KNEE FEM S LFT; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 6481-1-110
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code Ambulation or Postural Difficulties (2544)
Date of Event 07/22/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Event or Problem Description
The following was reported: femoral implant rupture at the junction between keel and femoral shield.Revision surgery required.
 
Additional Manufacturer Narrative
Reported event: an event regarding crack/fracture involving a mrh femoral component was reported.The event was confirmed via evaluation of the returned device and review of the medical records by a clinical consultant.Method & results: product evaluation and results: visual inspection: visual inspection of the returned devices was performed as part of the material analysis.On visual examination, the fracture was observed through the boss of the mrh knee, approximately 12 mm from the proximal tip.Material analysis: a material analysis was performed and concluded the following: "review of mrh knee fem xs lft, catalogue # 6481-1-110, lot code s3kdb confirmed a fracture through the boss.Characterisation using stereo microscopy and scanning electron microscopy confirmed the fracture propagated in fatigue." clinician review: a review of the provided medical information by a clinical consultant indicated: "the patients 2013 surgery was undertaken for chronic instability of a previously revised total stabilized knee.The patient apparently did well until the stem of the hinged femoral component fractured 8 years hence.In the post operative note there is commentary that the patient suffered a fall in 2020.It is unclear if this had any bearing on the stem fracture.The preoperative x-rays demonstrate a transverse fracture of the femoral stem at the level of the boss/stem interface.Following review of this documentation failure of the mrh femoral stem via fracture is confirmed as is the subsequent revision.The limited documentation precludes determination of a root cause for this failure." -product history review: review of the device history records indicate the devices were manufactured and accepted into final stock, with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to fracture of the mrh femoral component.The device was returned for evaluation and a material analysis was performed.The material analysis concluded the following: "review of mrh knee fem xs lft, catalogue # 6481-1-110, lot code s3kdb confirmed a fracture through the boss.Characterisation using stereo microscopy and scanning electron microscopy confirmed the fracture propagated in fatigue." a review of the medical records by a clinical consultant also confirmed the event and indicated that not enough information was provided to determine a root cause, although the reported patient fall may have contributed to the event.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
The following was reported: femoral implant rupture at the junction between keel and femoral shield.Revision surgery required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MRH KNEE FEM S LFT
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
EI   NA
Manufacturer Contact
anna ryan
raheen business park
limerick NA
EI   NA
61498200
MDR Report Key12506202
Report Number0002249697-2021-01595
Device Sequence Number4986462
Product Code KRO
UDI-Device Identifier07613327045130
UDI-Public07613327045130
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K002552
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2016
Device Model Number6481-1-110
Device Catalogue Number64811110
Device Lot NumberS3KDB
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/27/2021
Supplement Date Received by Manufacturer12/23/2022
Initial Report FDA Received Date09/21/2021
Supplement Report FDA Received Date01/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/07/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age71 YR
Patient SexUnknown
Patient Weight79 KG
-
-