Model Number 179712430
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Pain (1994)
Date of Event 08/24/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional product code: kwp, osh, nkb, mni, kwq.Complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2021 upon dissection to removed a broken medtronic rod, two screw head were detached.The screws were retrieved and replaced with medtronic screws.Procedure was completed successfully without any surgical delay.This report is for one (1) expedium spine system si polyaxial screw 5.5 x 4.35 x 30mm.This is report 1 of 2 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.A1, b5, d9, h3, h6: the device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was noted the patient reported pain and that was the reason for the revision procedure.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h4, h6: part # 179712440.Lot # apfdz9.The dhr of product code: 179712440, lot: apfdz9 was electronically reviewed and no non-conformances were observed during the manufacturing process.The product was released on: 02.05.2013.Qty: (b)(4).Visual inspection: the si polyaxl screw 4.35 x 40mm (p/n: 179712440, lot #: apfdz9) was returned and received at us customer quality (cq).Upon visual inspection, the locking screw had broken into two pieces at the point where the shaft meets the head of the screw.The screw was also stripped.No other issues were identified on the returned device.Device failure/defect identified? yes.Dimensional inspection: complaint relevant dimensional inspection cannot be performed due to the post-manufacturing damage.Document/specification review: the following current and manufactured revisions were reviewed.No design issues or discrepancies were identified.Complaint confirmed? yes.Investigation conclusion: the complaint could be confirmed for the returned device.No definitive root cause can be determined based on the provided information.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SI POLYAXL SCREW 4.35 X 40MM
Common Device Name
ORTHOSIS, SPONDYLOLISTHESIS SPINAL FIXATION
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
SENTIO (INNOVATIVE SURGICAL SOLUTIONS, LLC)
50461 west pontiac trail
wixom MI 48393
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key12506939
Report Number1526439-2021-02046
Device Sequence Number11577067
Product Code MNH
UDI-Device Identifier10705034067360
UDI-Public(01)10705034067360
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K160904
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number179712430
Device Catalogue Number179712440
Device Lot NumberAPFDZ9
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2021
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 08/24/2021
Supplement Date Received by Manufacturer08/31/2021
10/18/2021
Initial Report FDA Received Date09/21/2021
Supplement Report FDA Received Date09/23/2021
11/10/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/02/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SI POLYAXL SCREW 4.35 X 35MM; SI POLYAXL SCREW 4.35 X 35MM
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-