• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK 15 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK 15 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 15
Device Problem Failure to Sense (1559)
Patient Problem Cardiac Arrest (1762)
Event Date 08/23/2021
Event Type  malfunction  
Manufacturer Narrative
Physio-control contacted the customer to request additional information on the patient.The customer provided physio-control with the available patient information.Patient fields in which information is not provided were intentionally left blank.   physio-control evaluated the customers device and verified the reported issue.Physio replaced the device's therapy connector ribbon cable to resolve the reported issue.After completing other unrelated repairs, proper device operation was observed through functional and performance testing.The device was returned to the customer for use.
 
Event Description
The customer contacted physio-control to report that their device would not pick-up/read the pads.It would not let us select pads as a lead option as if they were not even connected.In this state the device may not be able to deliver defibrillation therapy if needed.This issue is patient related; however no health consequences or impact was reported.The customer indicated this was an asystole code and the pads were not needed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK 15 DEFIBRILLATOR/MONITOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key12507689
MDR Text Key272541363
Report Number0003015876-2021-01807
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00883873911631
UDI-Public00883873911631
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142430
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number15
Device Catalogue Number99577-001256
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2021
Initial Date Manufacturer Received 08/24/2021
Initial Date FDA Received09/21/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/16/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age34 YR
-
-