MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 20 DEG XLPE ACET LNR 36MM X 54MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 20 DEG XLPE ACET LNR 36MM X 54MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 71335754
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/31/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference case (b)(4).
 
Event or Problem Description
It was reported that, during a thr, a r3 20 deg xlpe acet lnr 36mm x 54mm had a black speck inside the plastic bag.This was noticed before use in patient.Surgery was performed, without any delay, with a smith and nephew back-up device instead.Patient was not injured as consequence of this problem.
 
Additional Manufacturer Narrative
Section: h3, h6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device did not confirm the stated failure mode.The device's packaging has a black spec within it.However, this spec is small enough to comply with the device's specifications.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use documents revealed this failure mode and potential harm was previously identified.Some potential probable causes could be but are not limited to packaging or manufacturing procedural/user error.Based on this investigation, the need for corrective action is not indicated.A field action review concluded that there are no actions related to this part number and failure mode.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 20 DEG XLPE ACET LNR 36MM X 54MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12509443
Report Number1020279-2021-07033
Device Sequence Number19676925
Product Code JDI
UDI-Device Identifier03596010598349
UDI-Public03596010598349
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K113848
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71335754
Device Catalogue Number71335754
Device Lot Number21BM13936
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2021
Supplement Date Received by Manufacturer12/10/2021
Initial Report FDA Received Date09/21/2021
Supplement Report FDA Received Date12/13/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-