|
Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this icd were re-investigated.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.As a first step in the analysis, the device was interrogated.The interrogation could be properly performed and it revealed the eri battery status.The device was implanted for approximately 90 months.Further analysis of the icd memory content showed a normal current consumption.However, the amount of charge taken from the battery was verified and the battery condition was not as expected.In order to obtain further insights, the icd was opened and the inner assembly was inspected.The visual inspection showed no anomalies.Subsequently the current consumption of the electronic module was verified by direct measurement and proved to be as expected and consistent with the interrogated icd data.There was no indication of a malfunction of the electronic module.Therefore, the battery was disconnected from the electronic module and sent to the manufacturer for further analysis.The manufacturing records of the battery were inspected, documenting that the battery parameters were within specification during the battery manufacturing.No anomalies were documented during the production process, associated with this battery.The battery was subjected to a visual and an electrical inspection as well as a microcalorimetry analysis.The visual inspection of the battery did not reveal any external signs of damage.The microcalorimetry analysis showed an elevated heat dissipation.In a next step, the battery was opened for destructive analysis.The inspection of the inner assembly identified damaged insulation, which led to the elevated internal self-depletion.Despite this finding, the battery was proven to have sufficient capacity to support therapy capability for at least the specified time from eri to the eos.In conclusion, the device was implanted for 90 months.The analysis revealed an elevated internal self-depletion within the battery.Based on the analysis all therapy functions of the device were entirely available while the device was implanted and in service.
|