• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG LIGATURE CLIP 12 MAG.= 144 PCS.; HANDHELD PRODUCTS & LIGATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG LIGATURE CLIP 12 MAG.= 144 PCS.; HANDHELD PRODUCTS & LIGATION Back to Search Results
Model Number PL572T
Medical Device Problem Code Unintended Ejection (1234)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/23/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.
 
Event or Problem Description
It was reported that there was an issue with pl572t - ligature clip 12 mag.= 144 pcs.According to the complaint description, it was reported that the magazine detached during the surgery.The product was replaced with a new one and could be used without any problems.There was no described patient harm.Additional information was not provided nor available / was not available.Additional information has been requested but not yet received as of this report.Additional patient information is not available.The adverse event / malfunction is filed under aag reference (b)(4).
 
Additional Manufacturer Narrative
Correction / additional information: d9: date returned to manufacturer.E1: region, telephone and fax.G2: distributor.G4: 510k number.H6: codes.Investigation results: visual investigation: the investigation was carried out visually and microscopically with the digital-camera.Here broken off noses, broken off latches and clip jams were found.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.The review of risk assessment revealed that the overall risk level (severity 3(5) x probability of occurrence 3(5)) according to din en iso 14971 is still acceptable.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.Based upon the investigations results a psc (b)(4) has been initiated.Any action regarding capa will be addressed with this case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGATURE CLIP 12 MAG.= 144 PCS.
Common Device Name
HANDHELD PRODUCTS & LIGATION
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key12510197
Report Number9610612-2021-00610
Device Sequence Number14739177
Product Code FZP
UDI-Device Identifier04038653341856
UDI-Public4038653341856
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K081031
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/26/2026
Device Model NumberPL572T
Device Catalogue NumberPL572T
Device Lot Number52679991
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2021
Supplement Date Received by Manufacturer04/04/2022
Initial Report FDA Received Date09/21/2021
Supplement Report FDA Received Date04/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-