| Model Number |
PL572T |
| Medical Device Problem Code |
Unintended Ejection (1234)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported that there was an issue with pl572t - ligature clip 12 mag.= 144 pcs.According to the complaint description, it was reported that the magazine detached during the surgery.The product was replaced with a new one and could be used without any problems.There was no described patient harm.Additional information was not provided nor available / was not available.Additional information has been requested but not yet received as of this report.Additional patient information is not available.The adverse event / malfunction is filed under aag reference (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Correction / additional information: d9: date returned to manufacturer.E1: region, telephone and fax.G2: distributor.G4: 510k number.H6: codes.Investigation results: visual investigation: the investigation was carried out visually and microscopically with the digital-camera.Here broken off noses, broken off latches and clip jams were found.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.The review of risk assessment revealed that the overall risk level (severity 3(5) x probability of occurrence 3(5)) according to din en iso 14971 is still acceptable.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.Based upon the investigations results a psc (b)(4) has been initiated.Any action regarding capa will be addressed with this case.
|
| |
|
Search Alerts/Recalls
|