| Model Number |
X060-1520 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The system implant plate is secured to the inserter/compressor by tightening a set screw on the instrument.It may be possible to fracture the distal tip of an inserter/compressor if excessive force was applied to the interface between the inserter/compressor and implant plate by over tightening the set screw.The root cause of this complaint cannot be reliably determined due to a lack of available information.The complaint investigation suggests the possibility that the set screw that secures the implant plate to the inserter/compressor was over tightened, resulting in the distal portion of the instrument breaking.
|
| |
|
Event or Problem Description
|
|
The company received notification on (b)(6) 2021 of a system inserter/compressor that was reported to malfunction when loading an implant onto the instrument.Five attempts to gather relevant complaint incident information and request the instrument to be returned were made, and no additional information was provided.There were no known patient complications reported as part of this instrument complaint.The instrument has not been received at the manufacturer for complaint assessment.
|
| |
|