• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 20 DEG XLPE ACET LNR 36MM X 56MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 20 DEG XLPE ACET LNR 36MM X 56MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 71335756
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994); Impaired Healing (2378)
Date of Event 08/31/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference case (b)(4).
 
Event or Problem Description
It was reported that, after a thr surgery, one syn por fem comp sz 13, one r3 3 hole acet shell 56mm, one oxinium fem hd 12/14 36 mm -3, and one r3 20 deg xlpe acet lnr 36mm x 56mm were associated with a post-operative reaction (pain, infection and unhealed incision).This happened after use in patient.The issue was resolved with a revision surgery, in which the surgeon removed all the implants.Current health status of the patient is unknown.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, this case reports the patient complaint of pain, infection and unhealed incision after a thr surgery.A ¿revision¿ surgery was performed to explant all devices.The patient¿s current health status is unknown.No relevant supporting clinical information has been provided to assist with a clinical investigation.The root cause of the revision/component explants was most likely the reported infection; however, the root cause of the infection, pain and unhealed incision could not be definitively concluded.The patient impact beyond the reported events could not be determined.No further medical assessment is warranted.A complaint history review found related failures for the listed device; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but are not limited to traumatic injury, joint tightness, patient reaction, contamination, patient reaction, and post-operative healing issue.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 20 DEG XLPE ACET LNR 36MM X 56MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12511046
Report Number1020279-2021-07042
Device Sequence Number3908311
Product Code JDI
UDI-Device Identifier03596010598356
UDI-Public03596010598356
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K113848
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71335756
Device Catalogue Number71335756
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/31/2021
Supplement Date Received by Manufacturer10/19/2021
Initial Report FDA Received Date09/21/2021
Supplement Report FDA Received Date10/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other; Required Intervention;
Patient Age65 YR
-
-