Model Number 74021165
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 06/19/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that, after tka surgery had been performed on (b)(6) 2017, the patient experienced a patella instability in the left knee.This adverse event was solved by revision surgery on (b)(6) 2018.Patient outcome as been resolved.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, per complaint details, the patient underwent revision due to left knee instability and pain approximately 12.5 months post implantation (ae log line#7).It was communicated that the patient status/outcome was ¿resolved¿.Based on the information provided in the crfs, the root cause of the reported events could not be definitively concluded.The patient impact beyond the reported instability and pain could not be determined; however, per correspondence, the outcome was ¿resolved¿.No further medical assessment is warranted at this time.Should additional clinically relevant documentation/information become available, the clinical/medical task may be re-evaluated.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to fit/sizing issue, lifetime of device or patient condition.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II CR FEM OX NP LT SZ 5
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED,
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12511127
Report Number1020279-2021-07056
Device Sequence Number3908319
Product Code JWH
UDI-Device Identifier00885556422786
UDI-Public00885556422786
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K121443
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number74021165
Device Catalogue Number74021165
Device Lot Number17EM19959
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/31/2021
Supplement Date Received by Manufacturer10/27/2021
Initial Report FDA Received Date09/22/2021
Supplement Report FDA Received Date10/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-