Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported material protrusion and fluid leak issues as no objective evidence was provided for review.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.(expiry date: 02/2023).
|
|
It was reported that one year post port placement, allegedly there was a popping sound during flushing the port.The patient allegedly had neck and arm swollen and sore.Upon x-ray examination, it was observed that the device allegedly had fluid leak.Patient current status is unknown.
|