MICRO THERAPEUTICS, INC. DBA EV3 AXIUM FRAMING COIL; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
|
Back to Search Results |
|
| Model Number |
FC-5-10-3D |
| Medical Device Problem Code |
Separation Failure (2547)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/17/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report that after coil deployment, the detachment was attempted but failed.Then, the manual detachment was attempted, however, this failed as well.The patient was undergoing surgery for treatment of a saccular, unruptured aneurysm in the left middle cerebral artery bifurcation (mcab) with a max diameter of 5mm and a 3mm neck diameter.It was noted the patient's blood flow was normal and vessel tortuosity was moderate.It was reported that after the coil deployment, the detachment was attempted but failed.Then, the manual detachment was attempted, however, this failed as well.The physician attempted to detach the coil with the instant detacher two times.A different instant detacher was not attempted.The manual attempt at detachment via the hypotube was attempted two times, as well.There was no issue determined with the instant detacher.The following coil was able to be detached with the instant detacher, which resolved the event.The product was replaced with a medtronic product.The device was prepared per the ifu.It was unknown if there were any patient symptoms or complications associated with the event.Ancillary devices include a 6f envoy microcatheter, a sl-10 guidewire, and a synco14.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information received reported that there were no problems prior to the non-detachment, and no damage was observed to the pushwire was observed after removal.
|
| |
|
Search Alerts/Recalls
|
|
|