Model Number MF-410375
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 06/14/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iii.
 
Event or Problem Description
Healthcare professional reported left side capsular contracture baker grade iii.Device remains implanted.
 
Additional Manufacturer Narrative
Device evaluation: analysis of the returned device identified the weight the device within specification, deformation, white particles in the shell and crease fold.Cloudy in the gel observed after autoclave cycle.Based on the device analysis the final assessment is: no issues found related with the manufacturing process.
 
Event or Problem Description
Device has been explanted and replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
chris sprague
12331-a riata trace parkway
building 3
austin, TX 78727
8479375615
MDR Report Key12512345
Report Number9617229-2021-52487
Device Sequence Number17429822
Product Code FTR
UDI-Device Identifier10888628003477
UDI-Public10888628003477
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P040046
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/15/2019
Device Model NumberMF-410375
Device Catalogue NumberMF-410375
Device Lot Number2683235
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/23/2021
Supplement Date Received by Manufacturer11/01/2021
Initial Report FDA Received Date09/22/2021
Supplement Report FDA Received Date06/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/15/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age55 YR
Patient SexFemale
Patient Weight75 KG
-
-