| Model Number |
PFM2CT5DL |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/12/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A dhr review was unable to be performed as the lot number was not provided.The risk management file was reviewed and includes this risk.Additional complaints have been rececived from this same customer.This appears to be a customer specific issue as no other complaints for the same or similar products have been received.The returned product was inspected and the cracked luer was observed.A different needle free valve was returned than the one provided.The possible cause of the issue could be the different needle free valve.It is recommended that the included needle free valve be used.Unable to determine the root cause.
|
| |
|
Event or Problem Description
|
|
Picc luer hub cracked.There was no injury noted to the patient.The picc catheter was removed.At this point, a peripheral catheter was placed and we will continue to determine if central access is/is not necessary.
|
| |
|