MAUDE Adverse Event Report: SMITH & NEPHEW, INC. UNKN ECHELON IMPL; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. UNKN ECHELON IMPL; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Catalog Number UNKN01100801
Device Problem Break (1069)
Patient Problems Fall (1848); Failure of Implant (1924)
Event Date 08/31/2021
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference case (b)(4).
 
Event Description
It was reported that, that a thr surgery was performed on (b)(6) 2021.The patient fell and broke the unkn echelon impl.This adverse event was solved by a revision surgery on (b)(6) 2021.Current health status is unknown.
 
Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, the broken echelon implant and subsequent revision is related to the patient fall on the date of implantation and is not associated with a malperformance of the implant.The patient impact beyond the revision and expected transient post-op convalescence period cannot be determined.No further clinical assessment is warranted at this time.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include but not limited to traumatic injury.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Additional information: d3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKN ECHELON IMPL
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12513314
MDR Text Key272711829
Report Number1020279-2021-07074
Device Sequence Number1
Product Code JDG
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 10/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/22/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKN01100801
Was Device Available for Evaluation? No
Date Manufacturer Received10/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age69 YR
Patient Weight93
-
-