| Model Number |
10607 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Used (b)(6) 2021 as event date since the correct date was not provided.Device evaluated by mfr.: synergy ii us mr 3.00 x 38mm stent delivery system was returned for analysis.A visual examination of the stent found stent damage.Stent struts on the proximal end were lifted and pulled distally.The undamaged stent outer diameter was measured and the result was within max crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of distal tip damage.A visual and tactile examination of the hypotube shaft found multiple kinks.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues along the shaft polymer extrusion.Device was tacked over a test 0.014 inch guidewire without issue.No other issues were identified during analysis.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 06sep2021.It was reported that loading difficulty was encountered.A 3.00 x 38 synergy ii drug-eluting stent was pulled out of the package and prepped in normal fashion.However, the physician was unable to load the stent on the wire.The procedure was completed with another of the same device.There were no patient complications reported.However, returned device analysis revealed a stent damage.
|
| |
|
Search Alerts/Recalls
|