STRYKER ENDOSCOPY-SAN JOSE PKG., AGG. PLUS SHAVER BLADE, F- SERIES - 4.0MM; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
|
Back to Search Results |
|
Catalog Number 0375544000 |
Device Problems
Break (1069); Entrapment of Device (1212); Appropriate Term/Code Not Available (3191)
|
Patient Problems
Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/26/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Additional information will be provided once the investigation has been completed.
|
|
Event Description
|
It was reported that there was a second procedure to remove the tip of the shaver from the patient.
|
|
Manufacturer Narrative
|
Alleged failure: customer reported that the shaver insert was broken.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be the blade comes in contact with a hard object causing damage to the teeth and hitting the housing edges.Excessive pressure, such as bending or prying, may cause the arthroscopic shaver blades to bend/break.Inadequate window profile.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
|
Event Description
|
It was reported that there was a second procedure to remove the tip of the shaver from the patient.
|
|