STRYKER ENDOSCOPY-SAN JOSE PKG., AGG. PLUS SHAVER BLADE, F- SERIES - 4.0MM; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL Back to Search Results
Catalog Number 0375544000
Device Problems Break (1069); Entrapment of Device (1212); Appropriate Term/Code Not Available (3191)
Patient Problems Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/26/2021
Event Type  Injury  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event Description
It was reported that there was a second procedure to remove the tip of the shaver from the patient.
 
Manufacturer Narrative
Alleged failure: customer reported that the shaver insert was broken.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be the blade comes in contact with a hard object causing damage to the teeth and hitting the housing edges.Excessive pressure, such as bending or prying, may cause the arthroscopic shaver blades to bend/break.Inadequate window profile.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
 
Event Description
It was reported that there was a second procedure to remove the tip of the shaver from the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG., AGG. PLUS SHAVER BLADE, F- SERIES - 4.0MM
Type of Device
BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key12514306
MDR Text Key272752481
Report Number0002936485-2021-00513
Device Sequence Number1
Product Code GFA
UDI-Device Identifier37613327061780
UDI-Public37613327061780
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 10/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/22/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0375544000
Device Lot Number21039CE2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2021
Date Manufacturer Received08/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-