• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. ALTIVATE REVERSE SHOULDER; ALTIVATE REVERSE, HUMERAL STEM ,STANDARD SHELL, SZ 8X108MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. ALTIVATE REVERSE SHOULDER; ALTIVATE REVERSE, HUMERAL STEM ,STANDARD SHELL, SZ 8X108MM Back to Search Results
Model Number 530-08-108
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 09/15/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was reported as loosening.The previous surgery and the surgery detailed in this event occurred 2 years and 8 months apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were kept at hospital and not made available to djo surgical for examination.This investigation is limited in scope as only partial information was provided to djo surgical for review.The revised items were not returned for examination and the lot numbers were not provided.To adequately investigate this event, the lot numbers are necessary.Given the limited information, a search of djo and available zimmer biomet records for the previous surgery produced no results, therefore; the items could not be verified.If this information is submitted at a future date, this investigation will be re-evaluated.Customer complaint history of the reported items showed no present trends or on-going issues that are needing a review.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.The surgeon performed this revision to remedy the patient's condition.This complaint will be closed pending receipt of additional information.No further action is deemed necessary at this time.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event or Problem Description
Revision surgery - patient's stem was loose.Osteolysis was present and cement was eroded.No infection was found.Baseplate/glenosphere was left in patient.Stem/poly/spacer was replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTIVATE REVERSE SHOULDER
Common Device Name
ALTIVATE REVERSE, HUMERAL STEM ,STANDARD SHELL, SZ 8X108MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
MDR Report Key12514934
Report Number1644408-2021-01026
Device Sequence Number3687204
Product Code PHX
UDI-Device Identifier00888912168304
UDI-Public(01)00888912168304
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K141990
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number530-08-108
Device Catalogue Number530-08-108
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/15/2021
Initial Report FDA Received Date09/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-