Model Number 0684-00-0295-05
Device Problems Filling Problem (1233); Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/15/2021
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation. Analysis of production - the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number. Based on the review results, it was determined that there is no relation between the manufacturing process and the reported failure. Historical data analysis - the review of the historical data was performed. Trend analysis- the overall complaint trend data for the period aug-19 to jul-21 was reviewed. Communication/interviews: communication/interviews were performed to obtain all possible information. Reference complaint #(b)(4).
 
Event Description
It was reported that during intra-aortic balloon (iab) therapy, the intra-aortic balloon pump (iabp) alarmed "autofill failure" and blood was found in the tubing. The iab was removed. There was no reported injury to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMEGA 7.5FR. 40CC IAB
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer (Section G)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer Contact
brian schaeffer
15 law drive
fairfield, NJ 
MDR Report Key12516431
MDR Text Key272891557
Report Number2248146-2021-00621
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K120868
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial
Report Date 09/22/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received09/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number0684-00-0295-05
Device Catalogue Number0684-00-0293
Device Lot Number3000128501
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/03/2021
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/14/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device?
Type of Device Usage Initial

-
-